Autoclave Validation and Performance Qualification

We can validate and perform Installation and Performance Qualifications of your Laboratory, Dental, Medical and Veterinary Autoclave in accordance with: 

 

AS/NZS 4815:2006 "Office-based health care facilities - Reprocessing of reusable medical and surgical instruments and equipment, and maintenance of the associated environment"  

 

AS/NZS 4187:2003 "Cleaning, disinfecting and sterilizing reusable medical and surgical instruments and equipment, and maintenance of associated environments in health care ”

  

The accuracy requirements of AS/NZS 4815 and 4187 are stated as follows.

 

The combination of sealed data logger plus indication/recording equipment shall be accurate to within ±0.5°C for temperature and have an accuracy of ±0.5% of the measurement span, or better for pressure.  

 

 

All measurements are traceable to national and international standards via a registered NATA laboratory or a mutually recognised international testing authority. (See NATA’s International Activities )