Autoclave Validation and Performance Qualification
We can validate
and perform Installation and Performance Qualifications of your Laboratory, Dental, Medical, Podiatry and
Veterinary Autoclave in accordance with:
AS/NZS 4815:2006
"Office-based health care facilities - Reprocessing of reusable medical and surgical instruments and
equipment, and maintenance of the associated environment"
AS/NZS 4187:2003
"Cleaning, disinfecting and sterilizing reusable medical and surgical instruments and equipment, and
maintenance of associated environments in health care ”
The accuracy
requirements of AS/NZS 4815 and
4187 are stated as
follows.
The combination of
sealed data logger plus indication/recording equipment shall be accurate to within ±0.5°C for temperature and
have an accuracy of ±0.5% of the measurement span, or better for pressure.
All measurements are traceable to national and international standards via a registered NATA
laboratory or a mutually recognised international testing authority. (See NATA’s International Activities )
|